The most commonly used capillary electrophoresis (CE) units are manufactured by Applied Biosystems. All Applied Biosystems CE instruments operate similarly. The two most common models used in forensic science are the ABI 310 and the ABI 3100. This section focuses on the ABI Prism® 310 Genetic Analyzer instrument. Other CE instruments used in forensic DNA laboratories are listed at the end of this module.

ABI Prism 310 Genetic Analyzer
National Institute of Justice (NIJ) (see reuse policy).

Autosampler Region

Component

Purpose

Autosampler

Holds the sample tray and tubes and presents them to the capillary for injection

Cathode Electrode

Provides a negative pole for electrical current and is composed of platinum

 

Autosampler Region
National Institute of Justice (NIJ) (see reuse policy).

 

Gel Block Region

Component

Purpose

Syringe

Stores the polymer between runs and generates the necessary force to fill the capillary with polymer

Anode Electrode

Provides a positive electrical current and is composed of platinum

Anode Buffer Reservoir

Contains buffer required for electrophoresis

Syringe Drive

Provides positive pressure to the syringe

Waste Vial

Collects waste

 

Gel Block Region
National Institute of Justice (NIJ) (see reuse policy).

Ions migrate through the capillary during electrophoresis. Positive ions will gather at the cathode and negatively charged ions will gather at the anode. The movement of ions creates an imbalance called buffer depletion. Buffer depletion can impair separation of DNA fragments due to a reduction in current. It is important to replenish or replace the buffer regularly to compensate for buffer depletion.

 

Detection Region

Component

Purpose

Charge-Coupled Device (CCD)

Detects fluorescence after the laser excites the dyes in the sample. The CCD camera is located behind the detector window.

Heat Plate

Heats the capillary during electrophoresis

Capillary

Where separation occurs. The capillary is a hollow fused silica tube with an internal diameter of 50-100µm and 25-75cm in length. There is an outer polyimide coating that provides physical strength while allowing a degree of flexing of the silica tube. The capillary is filled with polymer and is transparent to ultraviolet and visible light. The capillary itself can be used as the detector cell when fluorescence detection is used. Before each injection, the capillary is filled with a new aliquot of a viscous polymer solution.

Detection Cell Window

Where dye tagged product is detected. The detection cell window is produced by burning the polyimide outer coating off the capillary. This cleared area is directly in the line of the laser beam to detector path, allowing the DNA fragments to be visualized as they pass by.

Note: The detection cell window makes the capillary fragile in this area.

 

Detection Region
National Institute of Justice (NIJ) (see reuse policy).

 

Computer

Component

Purpose

Computer

Controls the operation of the instrument. The instrument settings are defined in method files in the computer. These settings consist of run temperature, voltage, injection parameters, rinse cycles, and sample positions.